mesoSPIM_excitation_path_galvo_rotation
The scanner should be rotated in its mount so that the beam is on axis if 0V are applied. When doing this step, is important to check again and again that the scanner is truly at 0V. Often, galvo scanners will move to a non-zero position when switched off.
Leveling the holder 30 that will carry the alignment collimator / laser.
Connect the Galvo scanner to a source with 0V / short the input to GND:
Put the alignment collimator into the holder 30 and stop the beam down to <10 mm diameter using an adjustable iris. Make sure that the beam hits the scan mirror in the center using the manual linear stage.
Attach a spare 60 mm cage plate and a few rods to be able to insert a 60 mm alignment target. Rotate the galvo scanner so that the beam is centered on the alignment target.
Depending on how you set up the system, do the same for the other excitation path:
Attach the cage plate to the scanner mount using two ER posts (each composed of an ER2 and an ER05). The posts are secured to the scanner mount using set-screws.
This is done using the digital indicator by indicating against the surface of the cage plate. Here, there is only 0.007 mm difference across the length of the cageplate - usually everything below 0.02 mm is ok.
-
Background
- mesoSPIM history
- Optical design
- Electronics
-
Setting up a mesoSPIM
- First steps
- Preparing the software and electronics
- Preparing the microscope optics
-
Setting the microscope up
- General alignment tips and tricks
- Installing the microscope base
- Setting up the detection path
- Alignment of the detection path
- Setup of the sample XYZ stages
- Setup of the excitation path
- Immersion cuvettes
- Set up a microscope config file
- Light-sheet co-alignment
- Set up initial ETL parameters
- Setting up lasers with the GUI
- Sample Handling
- Test Samples
- Troubleshooting
- Upgrades and custom variants