Readme Version: 1.1 (September 2017 / uploaded and formatted March 2021)
TPA Version: 3.3
Thanks for reading the TPA Readme Document!
Its purpose is to sum up the relevant information required to
run TPA smoothly and provide some guidance for troubleshooting.
This software originated during a DAAD RISE internship with Prof. Dr. Axel Zeitler's Terahertz Applications Group, Department of Chemical Engineering and Biotechnology, University of Cambridge.
During my terahertz spectroscopy measurements I thought about how the manual measurement process could be automated in a simple way, with the knowledge and tools available at that time. The result is my TPA software presented here. It is by no means perfect, and constitutes one of my first coding projects.
Note that the publication that was co-authored as part of the internship does not relate to the software uploaded here.
TPA automates all manual steps related to the measurement process, including
- moving a sample holder with two samples via a motor, manually done via the so-called IMS Terminal
- cooling the sample holder down with liquid nitrogen, manually by adjusting nitrogen flow
- heating the sample holder up to a specified temperature using a heating system
- checking and adapting the temperature using a temperature controller
- entering parameters and operating the TeraView software
This is done by
- controlling motor movement via a serial connection and PySerial
- operating the temperature control via a GPIB IEEE-488 connection and PyVISA
- executing mouse clicks and keyboard input via PyAutoGUI and the msvcrt package
This software comes without any warranty. It is not associated by any means with TeraView Ltd. or any of its products, such as TeraPulse.
THE SOFTWARE IS PROVIDED "AS IS", WITHOUT WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, TITLE AND NON-INFRINGEMENT. IN NO EVENT SHALL THE COPYRIGHT HOLDERS OR ANYONE DISTRIBUTING THE SOFTWARE BE LIABLE FOR ANY DAMAGES OR OTHER LIABILITY, WHETHER IN CONTRACT, TORT OR OTHERWISE, ARISING FROM, OUT OF OR IN CONNECTION WITH THE SOFTWARE OR THE USE OR OTHER DEALINGS IN THE SOFTWARE.
TPA is designed for the following hardware and software specifications:
-
Operating System: Windows 7
-
TeraPulse Software: Version 0.10.0 bb270313abf
-
TeraPulse Firmware: Version 0.5.2
-
Motor: MDrive 23
Intelligent Motion Systems, Inc. -
Temperature Controller: Model 331 Temperature Controller
Lake Shore Cryotronics, Inc. -
and suitable connectors for these components
To be able to run TPA, make sure to...
-
Connect all cables correctly:
- GPIB IEEE-488 cable for Lake Shore 331
- Power cable for MDrive 23
- Serial connection for MDrive 23
and turn both MDrive 23 and Lake Shore 331 on
-
Set a screen resolution of 1920 x 1080
The positioning system is based on this resolution, and any other resolution will lead to mouse clicks and inputs at wrong positions. -
Install Python 2.x (for
TPA32.py
) or Python 3.x (forTPA33.py
) and required packages- Install the PySerial (
serial
) package for handling serial controllers - Install the PyAutoGUI (
pyautogui
) package for automating keyboard and mouse controls - Install the PyVISA (
visa
) package for measurement device control - Install the Colorama (
colorama
) package for making ANSI escape character sequences work under Microsoft Windows - Install
matplotlib
andnumpy
for data processing and visualization - The
msvcrt
package is required to access routines from the Microsoft VC++ runtime (there is no multiplatform solution available unfortunately)
- Install the PySerial (
-
Enter correct GPIB settings for the Lake Shore 331:
- Baud Rate: 9600
- IEEE Address: 12
- IEEE Term: Cr Lf
- Emulation Mode: 331
It is of great importance to the proper execution of the TPA programme that the following steps are taken:
-
Open the IMS Terminal and load the preference
SingleMotor_Mike.ltm
. -
INITIALIZE the motor by typing the following parameters in the console. Press Enter after entering one parameter:
VA SP= 50000
RC= 100
VI= 500
VM= 40000
A= 30000
D= 20000
-
CENTRE the reference and sample holes properly.
Choose the correct centre position for both reference and sample holes and note them. -
Move the platform to the REFERENCE HOLE to the position you obtained in step III and set this position to ZERO, by typing the command
P = 0
. -
Now you may close the IMS Terminal to make the port accessible for TPA.
You will need the centre position of the sample for the initialisation process.
CAUTION: IT IS CRUCIAL THAT THE MOTOR HAS BEEN INITIALIZED, SET MANUALLY TO THE CENTRE OF THE REFERENCE USING THE IMS TERMINAL AND SET TO ZERO USING THE "P = 0" COMMAND. FAILING TO DO SO MIGHT END IN MOTOR FAILURE!
-
Open the command prompt
cmd.exe
You can access the command prompt by searching "cmd" in the search bar, when clicking on the Windows "Start" button, and then press Enter or follow this path:
C:\Windows\System32\cmd.exe
or simply doubleclick on theTPAxx.py
file. -
Move to the Desktop folder
Type
cd Desktop
to move to change directory.
The TPA file should be located in the Desktop folder.
If the TPA file has been moved, you should check the location of the TPA file and move to this directory using the "cd" command. -
Run the TPA file with Python
Enter
python TPAxx.py
to run the code. You should then see the starting screen, where you will be guided through the initialisation set-up.
-
TPA takes quite long to start a measurement/ TPA does not stabilise the temperature/ TPA does not start measuring at all:
If the temperature is not stable enough, TPA will not start a measurement.
If the tolerance is set too low (0.5K), the temperature controller might not reach a temperature within the allowed tolerance. Choose a tolerance of around 5K.
If the PID parameter are not correctly tuned, the waiting time might be too high for TPA. After the setpoint has changed, TPA waits for a time which equals the "P" parameter in [s]. Thus, if "P" is set too high, the waiting time increases.
If the flow of liquid nitrogen is not stable enough, frequent violations of the temperature tolerance might occur. To account for this, you could change the "D" parameter (press "PID/MHP" on the Lake Shore 331) and increase it. The controller will tackle sudden changes in temperature better.
-
TPA makes the temperature controller beep loudly:
The temperature controller beeps, if the current temperature is close to the set tolerance. Adjust your nitrogen flow in this case and try to achieve a stable flow.
If you hear a short beep followed by a long beep, this means that the tolerance has been exceeded and the measurement for the current temperature was terminated. Try to stabilise the temperature by adjusting the nitrogen flow.
You may open the console now and restart the measurement by following the steps indicated in the console.
-
TPA makes the motor move extremely fast/ TPA makes the movable platform crash to its upper/lower limit:
If this happens, the motor has not been initialised properly. Before you start running TPA, always open the IMS Terminal, initialise the motor with the commands listed in the "Before using TPA" section and make sure the motor is set to the centre of the reference hole and set to zero using the "p=0" command.
-
TPA makes my mouse move in a weird way while I use the computer/ TPA deletes things I typed and overwrites them:
Using the computer while the TPA programme is running is not possible, as it performs mouse clicks and keyboard inputs. You may only open the console window if a measurement has to be terminated when the set tolerance is exceeded.
Make sure that the TeraPulse window is opened, a new TeraPulse project file has been created and do not run any other programme that might pop up and jump in the foreground.
With software updates for the TeraPulse software to come, the interface might change.
Use the software with absolute caution after a software update has been performed!
To keep TPA working after a software update, conduct these steps:
-
Take a screenshot of the new TeraPulse interface.
-
Use an image editing software such as "paint.net" to obtain the new coordinates in [px] of the positions of all buttons that have to be clicked.
-
Open the TPA file with an editor such as "Notepad++" and scroll to the "INITIAL VALUES" section of the code
-
The locations of the buttons to press are stored in tuples, such as (354, 897).
-
Replace the positions with the ones you obtained using the image editing programme.
-
TPA will now move the mouse cursor to the correct positions of the buttons.
Please get in touch with me if you would like to get any information about TPA or support with troubleshooting!
I will be happy to help.