spatial-calibration

Python code to check the calibration of displays

This will involve displaying a bunch of images on your display, taking pictures of them with a high-quality camera (turning all the bells and whistles off, ideally storing as RAW images), and then running them through the included analysis to determine the displayed contrast.

Requirements

All python requirements are handled via the environment.yml file, but you will also need dcraw and exiftool installed (both are available in the default Ubuntu repositories and look easy to install on OSX as well).

The folder software_versions contains logs showing the versions used of dcraw, exiftool, gphoto2, and the python packages, as well as a simple python script used to ease the collection of images with gphoto2. Unless otherwise specified in the environment.yml file, it is not thought that the specific version matters for the gathering of the data, running of the analysis, or the analysis results.

All analyses only run on Linux (Ubuntu 18.04 LTS and CentOS Linux 7), but they should work with little to no tweaking on Mac. No guarantee they will work on any Microsoft OS.

If you are unfamiliar with setting up python environments, install miniconda for your OS, open your terminal, navigate to wherever you have downloaded this repo, and run the following:

conda env -f environment.yml

This will create a virtual environment with the necessary python libraries. Once you activate the environment by running conda activate calibration, python will use the version installed in that environment and you'll have access to all the libraries.

There are two main ways of getting jupyter working so you can view the included notebooks:

  1. Install jupyter in this calibration environment:
conda activate calibration
conda install -c conda-forge jupyterlab

This is easy but, if you have multiple conda environments and want to use Jupyter notebooks in each of them, it will take up a lot of space.

  1. Use nb_conda_kernels:
# activate your 'base' environment, the default one created by miniconda
conda activate 
# install jupyter lab and nb_conda_kernels in your base environment
conda install -c conda-forge jupyterlab
conda install nb_conda_kernels
# install ipykernel in the calibration environment
conda install -n calibration ipykernel

This is a bit more complicated, but means you only have one installation of jupyter lab on your machine.

In either case, to open the notebooks, navigate to this directory on your terminal and activate the environment you install jupyter into (calibration for 1, base for 2), then run jupyter and open up the notebook. If you followed the second method, you should be prompted to select your kernel the first time you open a notebook: select the one named "calibration".

Data

If you wish to re-run the analysis using pictures taken of the projector in NYU's Prisma during summer 2018 (as used by the spatial frequency preferences project), then you can get the raw data from the OSF (prisma_raw_images.tar.gz). The OSF also contains mtf-spline.svg and mtf-spline.pkl, which give the relationship between the modulation transfer function and the display frequency, as described below, and mtf.csv, which contains the data used to generate those files, as well as the MTF that results from different ways of measuring the contrast, preprocessing the RAW images, and photos taken at different times.

If you use snakemake as described in the Use section, the data will be downloaded and arranged onto the path as necessary for the rest of the analysis.

Use

If you have set up the environment, as described above, and will be using snakemake, as described below, this will be handled automatically.

Open up Snakefile and edit the DATA_DIR variable to wherever you would like to store the data (by default, this is the data/ directory within this directory). Then, navigate to this directory in your terminal and run:

conda activate calibration
snakemake -n -prk mtf_spline

Assuming everything is correctly configured, snakemake should wait a while as it builds the analysis DAG, then it should print out the many steps necessary to create the final output (see the overview section and the included Jupyter notebook MTF.ipynb to understand what steps are taken). To actually run the command, remove the -n flag (and you may want to add -j N, where N is an integer, in order to parallelize the jobs when possible).

At the end of this process, DATA_DIR will contain mtf-spline.svg and mtf-spline.pkl. Open up the svg file to view the modulation transfer function as a function of the display frequency in cycles per pixel (on a log scale). mtf-spline.pkl is a pickled python function (an interpolation of the data, as shown by the orange line in the plot), which takes spatial frequencies and returns the MTF. To load it in, run:

import pickle
with open('path/to/mtf-spline.pkl', 'rb') as f:
    mtf = pickle.load(f)

A simple example use:

# some spatial frequencies
sfs = np.logspace(-8, -1, base=2)
mtf(sfs)

See stimuli.py in the spatial frequency repo for a more involved example of how it can be used.

Overview

The goal of this project is to estimate the modulation transfer function (MTF) of a display, in order to examine how much contrast is loss at higher spatial frequencies. This happens because no lens (for a projector) or other display is perfect; all of them have a pointspread function of some form, which serves to blur (i.e., run a lowpass filter on) the displayed image, thus reducing the contrast at the higher frequencies. We want to estimate the approximate shape of this contrast vs. spatial frequency curve. As vision scientists, we will then use this information to construct stimuli that invert the MTF (reducing the amplitude of the low frequencies so that, when displayed, all frequencies have the same contrast), in order to evaluate how large of an effect this loss of contrast is.

In order to estimate the modulation transfer function, we created a series of square-wave gratings in a square aperture, each with a total width of 256 pixels, running from 1 cycle per image (256 pixels per cycle) up to the Nyquist limit, 128 cycles per image (2 pixels per cycle). We do this for both vertical and horizontal gratings. Each of these is surrounded by a white border and a black border, each with a width of 128 pixels, so that our overall displayed image fills a 512 by 512 pixel square in the center. These are displayed using PsychoPy, which enables us to perform luminance correction (so that a pixel whose value is 255 is twice as bright as one whose value is 127). We then take pictures of these images using a DSLR camera (we used a Nikon D90, but any DSLR should work; we used a standard consumer lens). The ISO was set to 200, the lowest available (in order to reduce sensor noise); the f-number (aperture size) was set to f16 (this was chosen so that the f-number was reasonably high / the aperture was reasonably small, in order to avoid lens artifacts, but not too high / too small, in order to avoid diffraction artifacts); the shutter speed was then set to ensure a reasonable amount of light reached the sensor (i.e., that the max recorded sensor value was about 80% of the max possible), while being long enough to avoid temporal aliasing from the refresh rate of the display and short enough to avoid shaking from small vibrations (the camera was set up on a linoleum floor on the ground floor of a building in New York University, directly above some subway tunnels). The camera was set up on a tripod, placed as close to the display as possible (to avoid any zoom) so that the full 512 pixels of the displayed image filled the picture, and connected via USB cord to a laptop running Ubuntu 18.04 (this tethering enabled us to capture images using gphoto2 (installed from Ubuntu package directory) without touching the camera, minimizing differences from picture-to-picture). The camera was focused manually as best as possible and photos were taken at two different focus settings to ensure they were approximately correct. The pictures were in raw image format (Nikon's variant is NEF) in order to avoid any post-processing performed by the camera (which could, for example, attempt to sharpen the image).

After the photos were taken, they were transferred to a laptop (the one running Ubuntu 18.04 mentioned above) and processed with the code found in this repository on some combination of the laptop, the NYU High Performance Computing cluster, and the Linux machines found in the NYU Lab for Computational Vision. There are two separate analyses, "first pass" and "mtf". Both of them start with demosaicing the raw image, which converts the images from raw sensory values into RGB or greyscale images. In order to make sure that this didn't have a large effect on the outcome, three different demosaicing methods were tried: a naive block-average method (average together the camera sensors in a 2x2 block so that you end up with a down-sampled greyscale image), dcraw's adaptive homogeneity-directed (AHD) interpolation, and dcraw's variable number of gradients (VNG) interpolation. This did not have any effect on the analysis outcome. An important step of both analyses is also the extraction of the grating found in the center of the image. Instead of relying on computer vision methods, such as edge-detection, we do this by simply finding the locations of the edges of the grating, the white border, and the black border (in pixels), recording them in pyspacal/camera_data.py, and using scipy.optimize.least_squares to simultaneously find the center of the square, the width of the grating, and the width of the border (both white and black borders have the same width) that best satisfies these points. The resulting square is used to construct masks corresponding to the grating and the two borders.

First pass:

  1. If the image is not already greyscale, average across the RGB dimension in order to convert it to greyscale.
  2. Extract the grating, as described above.
  3. Move a box across the grating, calculating the mean and standard deviation of the pixel values within the box. The box and step size should be either the same size as the grating's period or an integer multiple of it.
  4. Display these images and investigate. The mean image allows us to investigate how the average luminance changes, while the standard deviation image allows us to investigate how the contrast changes. Both can be viewed as a function of space (i.e., how they change within an image) and as a function of spatial frequency (i.e., how they change across images).
  5. We see that there is a hot-spot (the mean luminance is not constant across the image) and that the standard deviation decreases as spatial frequency increases. Mean luminance does not change as a function of spatial frequency, and the standard deviation has the same pattern of difference across an image (since the contrast is proportional to the mean divided by the standard deviation, the contrast does not change as a function of space)

MTF:

  1. We either convert the image directly to greyscale (averaging across the RGB dimension), or we first convert it to a luminance image, following the procedure in Tkacik et al, 2011:
  • Standardize the RGB values. Tkacik et al show their camera sensor values are linear with respect to shutter speed, aperture, and ISO, and we assume this; we do not check this.
  • Convert the RGB values to LMS (cones), using the sensitivities of the Nikon D90 camera from the camspec database (which give the responses of the R, G, and B sensors as a function of wavelength) and the 2 degree cone fundamentals (which give the responses of the L, M, and S cones as a function of wavelength).
  • Convert LMS to luminance using the values given in the Tkacik et al paper.
  1. Divide the greyscale or luminance image by the greyscale or luminance version of an image taken of a mid-grey stimulus on the same display, taken during the same session with the same settings. This will effectively correct for the hot-spot problem mentioned above, leaving only the variation in contrast as a function of spatial frequency.

  2. Extract the grating, as described above.

  3. Calculate the contrast of the grating, $I$, using one of three contrast measures:

    • Michelson contrast: $\frac{\max(I) - \min(I)}{\max(I) + \min(I)}$
    • Root mean squared contrast: $\frac{\sqrt{\frac{\sum{I_i - \bar{I}}}{n}}}{\bar{I}}$, where $I_i$ is the value of pixel $i$ within the the grating $I$.
    • Fourier contrast: convert the 2d grating into a 1d grating by averaging along either the vertical or horizontal direction (depending on whether the grating was horizontal or vertical), and take the amplitude of the square wave's fundamental (we do not calculate the frequency of the fundamental, we simply take the Fourier transform of the 1d grating and find the frequency with the max amplitude).
  4. Plot the contrast of each grating as a function of its frequency (which we know based on how we constructed the grating).

  5. This plot shows the modulation transfer function of the display. For a given image, Michelson > RMS > Fourier. Because the display's pointspread function acts as a lowpass filter or blurring, the Michelson contrast will be least affected (the blur will have the greatest effect on the pixels near the transition of the square wave grating and will need to grow fairly large before affecting the extreme values). The RMS contrast is more affected by the blur than the Michelson contrast, but is affected by multiple frequencies, whereas the Fourier contrast comes from the fundamental and thus will be the lowest. The shape of the MTF curve is the same for all three contrast measures, but the actual contrast values differ. The use of luminance image or the greyscale image calculated directly from sensor values does not seem to have any effect on the measured MTF.